cooling of blood platelets clusters the von
Willebrand factor (vWF) receptor complex. Macrophage aMb2
integrins bind to the GPIb (a
+ b)
subunit of the clustered complex, resulting in rapid clearance of transfused,
cooled platelets. This precludes refrigeration of platelets for transfusion,
but the current practice of room temperature storage has major drawbacks.
We document that aMb2
is a lectin that recognizes exposed ß-N-acetylglucosamine residues
of N-linked glycans on GPIb. Enzymatic galactosylation of chilled platelets
blocks aMb2
recognition, prolonging the circulation of functional cooled platelets.
Platelet-associated galactosyltransferase produces efficient galactosylation
when UDP-galactose is added, affording a potentially simple method for
storing platelets in the cold. Transferring the technique to stem cells
is tricky because, unlike platelets, they have nuclei that must be perfectly
preserved.
Today researchers save stem cells by freezing them in expensive vats of
liquid nitrogen at -140 ºC. To stop damaging ice crystals forming
inside the cells, they bathe them in a toxic chemical such as DMSO.
This must be washed off before cells can be used.
Web resources : dmoz
Open Directory Project : Cryobiology
liophilization
/ cryodrying : high-speed freezing (to create only small-sized H2O
crystals) followed by transition to gaseous state. A buffer is needed to
mantain Aw = 1%.
air drying :
several scientists are looking for a way to air-dry cells, comparable to
the technique that turns grapes to raisins. Dried stem cells might be stored
in portable packets that can be ripped open and their contents mixed with
water at the scene of an accident. After drying and soaking the stem cells
in the anti-freeze sugar trehalose (which sloshes around in organisms
that survive dehydration, such as brine shrimp, baker's yeast and certain
drought-resistant resurrection plants) and arbutin, < 40% of
the cells was water. When rehydrated immediately afterwards, up to half
grew and divided and cells can be revived after several weeks on the shelf..
Dehydration must remove as much water as possible so that cells' metabolism
is halted, otherwise they churn out waste that poisons their surroundings.
But cell membranes crack if they are overly parched. Researchers are experimenting
to strike the right balance.
Purification
of cell types
tissue microdissection : dissection
of tissue or cells under the microscope.
laser-capture microdissection
(LCM) : energy from a low-power laser fitted to an inverted microscope
is used to melt a thin vinyl film to precise loctions on a tissue section
to bind targeted cells. After the appropriate cells have been selected,
the film and adherent cells are removed for gene-expression studies.
magnetic
activated cell sorter (MACS) : immunomagnetic microbeads [movie]
cytometry : the characterization and measurement
of cells and cellular constituents.
microfluorometry / cytophotometry
: the study of organic compounds within cells by means of the cytophotometer
(a photometer for measuring localization of organic compounds within cells
by measuring the light intensity through selected stained areas of cytoplasm)
image cytometry : a technique in which
histologically prepared cells are imaged using a scanning technique that
divides the whole image into many smaller elements and aspects, which can
then be analyzed by a computer and compared between many different cells
in a series.
flow cytometry
: a technique in which cells suspended in a fluid flow one at a time through
a focus of exciting light, which is scattered in patterns characteristic
to the cells and their components; they are often labeled with fluorescent
markers so that light is first absorbed and then emitted at altered frequencies.
A sensor detecting the scattered or emitted light measures the size and
molecular characteristics of individual cells; tens of thousands of cells
can be examined per minute and the data gathered are processed by computer.
The cell structure(s)/molecule(s) are marked with
a fluorophore (direct ligand)
enzyme substrate
nucleic acid binder
a fluorophore-conjugated ligand
monoclonal antibodies (immunophenotyping)
: pelletize antigen-bearing cells (5 x 106 each in 3 wells :
the first with no antibody added; the second with only secondary antibody
added; and the third with both immune serum and secondary antibody added)
by centrifuging 3 minutes at 1,600 rpm; add 100 mL
of FACS buffer (PBS 3%, 0.2 NaN3) per well and pelletize again;
add 1 mL/well of serum from surviving animals to the second and third wells
to obtain a final dilution = 1 : 200. Incubate for 20' over ice (covered
with aluminum paper) and pelletize again. Washing with 100 mL
of FACS buffer. Add 1 mL of secondary goat anti-mouse
mAb to the third well only. Incubate for 20 minutes and pelletize again.
Add 100 mL of FACS buffer and read at FACS (after
usage clean the FACS with hypochlorite first and water then)
lectins
cytokines
nucleic acid probe
beads or particles
... and the unbound is washed out (or, in the case of substrates, maintains
another excitation peak). All fluorophores
need to be stimulated from the same laser beam, and hence need a common
(or at least similar) excitation spectrum. Stokes's shift
: lexcitation - lemission
(nm).
A cytometer consists of :
optics
necessary for excitation
light source(s), usually laser(s)
argon ion lasers => 488 nm (the absorption peak of several commonly used
fluorochromes and dyes)
helium-neon lasers (He-Ne) lasers => 633 nm
helium-cadmium lasers (He-Cd) => 325 nm
red diode lasers => 635 nm
focusing lens
necessary for collection of the light emitted from, or scattered by, the
cells.
detection and quantitation of laser light scattered at
90° : orthogonal or side scatter (SS), correlated with the granularity
of the cell
180°C : forward scatter (FS), correlated with cell size
detection and quantitation of as amany as 11 colors of immunofluorescence
emanated at 90°
if the fluorescent light is separate as a result of sequential laser
intercepts, then the signals may be sent in dfferent directions through
the use of
beam splitter
half mirrors
if all fluorescent light is mixed
dichroic mirrors reflect light above a specified wavelength and
permit light below that wavelength to pass through
short pass filters absorb light above a specified wavelength and
permit only light below a specified wavelength to pass through
long pass filters absorb light below a specified wavelength and
permit only light above a specified wavelength to pass through
bandpass filters permit only light in a specified range of wavelenght
to pass through, absorbing light with wavelength above and below that range
488 BP filter
fluidics : coaxial stream whose pressure is controlled by electronics.
inner sample stream : at least 1,000 up to 100,000 cells per second
outer sheath stream to focus the inner stream and to minimize the
turbulence arising from the resistance of the tubing, allowing cells in
the sample stream to be presented to the laser one at a time and in a sequence
that can be maintained for interrogation by multiple lasers.
Although most clinical assays do not use sorting (fluorescence
activated cell sorter (FACS)), it is commonly used in research.
For sorting, the coaxial stream is broken into droplets after interrogation
by the laser, and these droplets are then charged according to the fluorescence
of the cell (or cells) contained within the droplet. The drops are then
sorted by charged deflection plates (controlled by electronics) that alter
the path of the drop as a result of the charge. Some cytometers sort cells
in 4 directions.
The liquid waste is transported by a vacuum into a tank which should
contain a disinfectant such as bleach to minimize biohazard risk and must
be manually emptied at periodic intervals.
electronics
photomultiplier tube (PMT) / photodiodes convert the incident light
into electronic pulses. The sensitivity of the PMTs to light is controlled
by the voltage supplied to the PMT, thus electronics are needed to control
this voltage
signal amplification
analogue-to-digital converters (ADCs) for computer analysis
compensation circuits controlled by electronics
computer (via Institute for Electronic and Electrical Engineers (IEEE)
port) : CellQuest™ software
controlling instrument operation
analysis of data
presentation of data
2-parameter histograms
dot plots
export statistics for use in spreadsheets, used for further analysis and
graphics
storage of data in listmode files (usually on CD-ROMs due to lack of high
capacity storage devices)
algorithms used to compare data
Kolmogorov-Smirnov
(K-S) two sample algorithm
is a commonly used method to determine the confidence interval (D value)
with which one can make the assertion that 2 flow cytometric univariate
histograms are different (P = 0.001)
Overton cumulative histogram subtraction algorithm essentially subtracts
histograms on a channel-by-channel basis to provide a percent of positive
cells.
super-enhanced Dmax subtraction (SED) is a new sophisticated algorithm
by Bruce Bagwell to compute %Positives when comparing histograms.
probability binning (PB) comparison is related to the Cox chi-square
approach, but with modified binning such that it minimizes the maximal
expected variance. This algorithm has been shown to detect small differences
between 2 populations and it does so in a quantitative way.
Single cell analysis : if clumps or doublets causing
misinterpretations (e.g. CD4+ cell + CD8+ cell =
DP cell ; 2 nuclei = G2/M cell) are known to exist in the specimen,
they can be filtered out before analysis or gated out using the appropriate
parameters.
Compensation to subtract the portion of the observed fluorescence
resulting from spectral overlap :
pre-data acquisition compensation
visually determined (and manually set)
established by software programs
post-data acquisition compensation
established by software programs
Sequential gating with Boolean logic :
based on light scatter (traditional SS-to-FS plot). E.g. for peripheral
blood :
based on immunofluorescence
... performed
during acquisition of data
during analysis of listmode files : the gating is not permanent and list
mode data may be regated an infinite number of times if the original file
is not overwritten
Quantitative cytometry :
express the detected fluorescence in molecules of equivalent soluble
fluorescein (MESF units)
quantitate antigen density in antibody binding capacity (ABCs) in
immunophenotyping
[cells having the target molecule]sample volume =
= [cells having the target molecule]%sample volume, detected
with cytometer. absolute number of cellssample
volume, performed manually or on other instrumentations
= [fluorescent beads]sample volume, known.
[fluorescent beads]%sample volume, detected with cytometer /
[cells having the target molecule]%sample volume, detected with cytometer.
This method has higher accuracy and precision.
Quality control :
of the flow cytometer : analyzing beads (or other particles) with known,
stable, fluorescence and light scattering properties
of the reagents and specimen preparation techniques : staining and analyzing
normal control specimens and comparing the data to a historical database
BD Biosciences, a segment of Becton, Dickinson and Company (BD), developed
one of the earliest commercial flow cytometers, called a fluorescence-activated
cell sorter (BD FACS), in the early 1970s. For this reason, some people
still use the term FACS generically to mean flow cytometry and the instruments
that perform it. Today FSC and SSC are taken at breathtaking speeds exceeding
70,000 cells per second with unsurpassed accuracy and sensitivity. For
this reason, flow cytometry is ideally suited for studies that need to
distinguish among up to several million cells based on many different parameters.
The recent marriage of flow cytometers with other detection systems, such
as imaging devices and multiplex bead-based assays, has greatly enhanced
the amount of information that can be gathered from each cell. A major
development in flow cytometry has been the availability of an arsenal of
fluorescent probes. In addition, it is now possible to more precisely quantify
the amount of signal coming from one cell relative to another. E.g. researchers
can identify receptors on the surface of cells, total DNA, the amount of
mRNA for a particular gene, calcium concentrations, membrane potential
and pH. Some of the newer probes allow researchers to measure the amount
of protein phosphorylation in a cell to study signaling cascades. BD Biosciences
alone sells approximately 25 types of fluorescent reporter dyes coupled
to antibodies, probes and other reagents. To take advantage of the growing
number of fluorescent reagents, flow cytometers have multiple lasers to
stimulate different excitation wavelengths and several detectors to collect
different color signals. The top-of-the-line machine produced by BD Biosciences
can detect up to 18 colors. Germany-based Partec, which started distributing
flow cytometers in 1969, offers different configurations of instruments
and lasers to suit customers' needs.There is very little limitation for
detecting more colors on the instrument side. But despite the boundless
possibilities, only a handful of published studies look at > 8 colors simultaneously
and most will use 4-5. Complex multicolor analyses are difficult to
conduct : one of the main stumbling blocks is the data processing and proper
interpretation
if you have ever had to do it, you know the trials and tribulations associated
with generating a stable cell line that expresses your favorite gene. Chromocell,
a startup company based in New Jersey, has found a way to deliver stable
cell lines that reproducibly and reliably express high levels of functional
proteins with greater ease and efficiency. The technology uses a combination
of special fluorescent probes to detect the mRNA corresponding to the gene
of interest in individual transfected cells and high-speed sorting of the
target cells by flow cytometry. When the probes are added to a batch of
cells, only those cells that contain the target message end up with a fluorescent
signal. The cytometer is then programmed to select cells with the highest
amount of signal so that they can be further studied to determine whether
the expressed protein is functional. For proteins made up of more than
one subunit, cells are transfected with, for example, 3 individual genes
and the cytometer simultaneously tests for the presence of all 3 signals
in a cell. The procedure is largely automated making it possible to screen
millions of cells in a matter of minutes. When you get so many cell lines
that express the gene you are interested in, you can then select the cells
that meet other criteria, for example stability. In addition to stability,
the researchers test the response to different treatments to see whether
the expressed protein behaves as it should. Another advantage of the technology
is that unlike high-throughput methods for cell selection in which transfected
genes carry some kind of fluorescent tag, such as a GFP molecule, the expressed
protein is as close to the native configuration as possible. The company,
which was established in 2002 by researchers from Günter Blobel's
lab at The Rockefeller University, produces stable cell lines as a service
and is now exploring other applications of the technology, in particular
for antibody biologics production
every flow cytometer includes software for collecting and analyzing information
gathered during an experiment. The analysis typically involves finding
out how many cells from the sampled population meet a criteria of interest,
and the data can be displayed in a number of different formats, most frequently
dot plots or histograms. Basically, histograms quantitate intensities of
scatter or fluorescence based on one parameter, whereas dot plots quantitate
percentages of cells with various properties. Dot plots are typically used
to detect small numbers of cells that are well separated from the main
populations of the cells present. In addition, in the software packages
that come with some imaging cytometers, such as Amnis' IDEAS
data analysis software, the dot plots are linked to images of the
cells. That means that a user can click on an individual dot to see
the corresponding image or can draw a 'gate' around a population of dots
to see what that cells that fall within the gate look like. This is very
useful for confirming that the dots really do represent the cells you think
they do and aren't clumps of cells or other artefacts. Many companies also
sell software packages separately from the instrumentation. One of the
most popular products is Tree Star's FlowJo
software, which can analyze data from any flow cytometer and will work
in both a Windows and MacIntosh environment. FlowJo is based around 'workspace'
documents—similar to pages in a notebook—that contain information about
the samples, along with all the gates, graphs, statistics, tables and reports
applied to the data. FlowJo provides several different analysis platforms
for standard analyses such as gating and statistics, as well as more special
ones, such as DNA content, cell cycle analysis, kinetics, proliferation,
calibration and statistical comparisons. Other examples include BD
CellQuest, a popular commercial flow cytometry data acquisition and
analysis software for the MacIntosh computer produced by BD Biosciences.
FSC
Express from De Novo Software is a very advanced analysis product that
runs on the PC. Verity has Winlist,
a comprehensive data analysis package for the PC, and Modfit, one of the
main DNA and cell cycle analysis packages. Some software packages are geared
toward more sophisticated applications. Applied Cytometry Systems (ACS)
has developed the StarStation
software to drive the Luminex flow
cytometer and help analyze the data generated from it. Bio-Rad and
MiraiBio have similar software products called Bio-Plex
Manager and MasterPlex,
respectively. The software takes care of calibrating the instrument, running
the assays and collecting the data. In September 2005, ACS will release
a new product that reads files generated in StarStation and analyses them
in custom ways according to a particular plug in application. The first
plug in, called SNP-R, will be used to compare a pair of alleles to determine
which one is present in a given sample. It is similar to data obtained
using a SNP microarray but different in that the sample is in flow. The
company plans to continue developing different types of plug-ins for customers.
Improved lasers have not only given way to greater accuracy and sensitivity,
but also smaller instruments. Lasers are becoming smaller and cheaper and
so are the instruments. In the past 3 decades, high-end machines have gotten
smaller without compromising power. DakoCytomation's CyAn
cytometer contains 3 lasers (blue, red and violet), and can measure
forward and side scatter and up to 9 colors of fluorescence, all at speeds
of over 30,000 cells per second. According to the company, the instrument
is 6-10 times smaller than earlier models with comparable capabilities.
BD Biosciences sells a low-cost benchtop analyzer called BD
FACSArray. With 2 lasers, it can measure 15,000 cells per second and
visualize 4 colors, as well as cell size and heterogeneity. In addition,
several companies, including Guava Technologies, Beckman Coulter and Partec,
sell small 2-color units that are more targeted to specific applications,
such as measuring cell proliferation or apoptosis. Customers like the fact
that they use their budget for the configurations that they need. The most
recent newcomer in this arena is Beckman
Coulter's Cell Lab Quanta flow cytometry system, which began shipping
in April of 2005. Built on a joint development effort with NPE Systems,
the Quanta combines Coulter volume measurement (an electrical sensing method
for counting and sizing cells), with 2-color fluorescence capabilities.
In September 2005, the company will release a new model that can also measure
light scatter and one additional color. The size and affordability of these
smaller instruments, as well as the ease of use, have made it possible
for them to be used in developing countries to monitor HIV infection in
the population.
flow cytometry was thrust into the spotlight by the AIDS epidemic in the
1980s as it provides a quick and accurate way to count CD4+
and CD8+ T cells in HIV-infected patients. Until recently flow
cytometers and the necessary reagents were too expensive and too difficult
to implement in developing countries—where HIV infection is ravaging the
population. Thus, the development of smaller and more affordable, robust
instruments has had a substantial impact in this part of the world.
Partec has 2 kinds of mobile flow cytometry units: one, housed in a car,
which is powered by the car battery and solar panels, and the other, part
of a laboratory truck, is powered by its own generator. Since 2002, > 200
CyFlow
instruments have been placed in Sub-Saharan Africa, offering CD4 counting
at a cost of about $2.2. In addition to CD4 counts, the Partec instruments
measure the number of CD4+ T cells as a % of the total number
of lymphocytes. Recent studies have revealed that in infants, CD4 counts
normally decrease with age, whereas the proportion of CD4 remains stable.
This means that the threshold of 200 CD4+ T cells generally
used to decide when to start antiviral therapy cannot be safely used for
children. The decision to start therapy in pediatric patients should instead
be based on the CD4 % value. Guava Technologies' EasyCD4
system weighs a mere 35 pounds. These systems and optimized CD4 assay
kits have been sold throughout Africa, India and southeast Asia for CD4
counting at slightly > $1 per test. Other companies offering flow cytometry
products for CD4 counting in developing countries include BD Biosciences
and Beckman Coulter.
Some flow cytometers can physically separate a group of cells from the
rest. The instrument puts a negative or positive charge on specific cells
that have defined properties, and then deflects these charged cells to
separate collection tubes. Because multiple fluorochromes can be assessed
simultaneously,
cell sorting by flow cytometry can separate complex mixtures of cells on
the basis of several markers—a capability that is increasingly exploited
in stem cell research protocols. As a recent example, Sean Morrison's group
at The University of Michigan, Ann Arbor, was able to isolate mouse HSCs
of unprecedented purity from blood, based on the combined expression of
the receptors of the SLAM familyref.
The ability to do these kinds of experiments rests partly on improved sensitivity
of detection. But another requirement for cell sorting, especially when
dealing with a population of rare cells, is speed. DakoCytomation's MoFlo
was the first true high-speed cell sorter and the first to allow separation
of four different cell populations simultaneously. It has high performance
in terms of yield, cell viability and purity, and runs at very high speed,
reaching up to 70,000 cells per second."The sorter has an open design,
which means that the user can add different kinds of lasers as new fluorechromes
are developed. Another popular cytometer with cell sorting capability is
BD
Biosciences' FACSVantage SE System. With a new generation of lasers
and advanced optics, the company claims it provides enormous flexibility
and sensitivity. It sorts up to 4 populations of cells simultaneously at
rate of 16,000 cells per second, with sorted fractions exhibiting > 99%
purity. Traditionally, flow cytometers with cell sorting capability were
large instruments, but that is no longer the case. In April 2003, BD Biosciences
began shipping the first high-performance benchtop cytometer with electrical
sorting capabilities and, according to the company, the instrument, called
BD
FACSAria, has taken the market by storm. Improvements in fluidics and
optics allow for higher speed sorting (25,000 events per second) with less
stress on the sorted cells and enhanced overall sensitivity. The system
uses 3 lasers to allow detection of 13 colors as well as scatter. An exciting
advance has been the combination of flow cytometry with imaging tools and
devices. Amnis Corporation pioneered multispectral image flow cytometry,
a method that allows individual cells to be visualized as they zoom by
the laser beam. Amnis' ImageStream
100 is similar to a flow cytometer with the important distinction that
the traditional detectors have been replaced by a sensitive CCD camera,
custom designed by the company. The camera takes as many as 6 independent
images (brightfield, darkfield and up to four fluorescent images) of each
cell at a rate of about 100 cells per second. Although the speed is slow
for a flow cytometer, it beats that of any high-content screening system
hands down. Because it can image a large number of cells at once, the ImageStream
is ideally suited to gather a statistically meaningful picture of a population
of cells. If there is any variability in a cellular response to a treatment,
you will not get the necessary statistical information using a microscope.
The instrument provides the same measurements as a traditional flow cytometer
but with the added benefits that you can also calculate where the fluorescent
signal is coming from in the cell. For example, a researcher can measure
the localization of 2 cell-surface markers relative to one another or determine
the cytoplasmic versus nuclear location of a molecule, all in a large population
of cells. Many researchers believe that tumors contain a rare population
of therapy-resistant resting cells, which act like stem cells to give rise
to cells that continuously divide. Laser scanning cytometers (LSC)
are also capable of visualizing the location of fluorescent signals in
a cell, but using a different strategy. In an LSC the sample does not move,
rather the laser beam sweeps over the cells to construct an image in a
manner similar to confocal microscopy. Unlike a microscope, the LSC allows
the user to scan a relatively large area without the need to refocus. In
addition, the LSC allows the quantification of all the fluorescent light
emitted from the entire cell depth at each location in the specimen, just
as a flow cytometer does. It has the same quantitative power as flow cytometry
in terms of sensitivity and precision, but with the addition of imaging
and the ability to work with live or fixed adherent cells, using a variety
of carrier types. CompuCyte's LSC
has 4 fluorescence and 2 scatter detectors, to capture up to 6 images of
cells; the sample can be either in suspension, adhering to a slide or in
tissue sections. The LSC is well-suited for applications that cannot be
done in flow. It is good for studying apoptosis in adherent cells. Removing
cells from a culture dish to suspend them can simulate cell death, so it
is preferable to analyze them while they are still on their growth surface.
CompuCyte's laser scanning cytometers can analyze both fluorescent and
chromatic dyes, as well as combinations of the 2. This extends cytometric
investigations to objective cytometric analysis of tissue sections without
compromising tissue architecture and allows pathologists to view samples
in the manner that they are accustomed to, coupled with additional quantitative
data similar to flow cytometry report. In 2006, the company will provide
an additional feature with their instruments to allow the measurement of
decay times of fluorescent labels. Because conditions like pH, oxygen or
cation concentrations affect the rate of decay of a fluorescent dye, decay
times can provide a sensitive measure of the characteristics of a cell's
microenvironment. Another LSC system, called SurroScan,
was developed at SurroMed Inc. (which was acquired in early 2005 by Pharmaceutical
Product Development Biomarker Discovery Sciences, a company that provides
cytometry services for clinical studies and trials). The system combines
the best features of a hematology analyzer and a flow cytometer. Samples
are placed in a volumetric capillary array that holds 32 assays with 3-4
colors in each. As the laser scans over each capillary, the instrument
gathers information to classify and quantify cells in blood or other fluids.
Results include both absolute cell counts and relative levels of antigen
expression. You can process a large number of assays on a large number
of samples in a reproducible way. New systems using multiplex bead arrays
push the potential applications for flow cytometry even further. These
systems simultaneously measure up to 100 compounds in suspension using
very small volumes. Common applications include profiling cytokine concentrations
in serum or identifying which proteins are phosphorylated after activation
of a given kinase. The flow cytometers and reagents
developed by Luminex combine the power of flow cytometry with the simplicity
of an ELISA plate reader. Instead of reading cell-based assays, they read
a microsphere-based assays. The Luminex method uses microsphere sets carrying
variable quantities of two different fluorescent dyes that produce up to
100 different shades of color. Each bead is coupled to a unique antibody
or probe that recognizes a specific molecule. After the beads are mixed
with a sample—for example, cell lysates—and added to the instrument, the
unique color signature on each bead reveals the identity of the bound molecules.
Several companies, including Applied Cytometry Systems, Biosource International,
MiraiBio, Qiagen, Upstate, LINCO Research, Marligen Biosciences and others,
are licensed to sell Luminex instruments or reagents. In addition, other
companies have developed similar technologies. BD Biosciences' FACSArray
analyzer can process bead-based assays using the company's own brand of
reagents. Beckman Coulter also produces multiplex bead assays that, instead
of having 2 colors on one bead, use 2 beads per assay. Although relatively
new on the market, the power of this technology is clear. Todd Golub and
Robert Horvitz used it to analyze the expression of microRNAs (miRNAs)
in different human cancersref.
As miRNAs are only about 21 nucleotides long and share similar sequencesref,
it is difficult to detect specific miRNAs using glass slide microarrays—a
popular method for expression profiling. To overcome the problem, the researchers
coupled oligonucleotide probes complementary to the miRNAs of interest
to fluorescent beads, whereby each bead represented a single miRNA. They
added the beads to miRNAs isolated from cancer cell lysates and stained
with streptavidin-phycoerythrin. When the samples were run through a flow
cytometer, the instrument simultaneously measured bead color (miRNA identity)
and phycoerythrin intensity (miRNA abundance) to reveal a different miRNA
expression profile for each sample. The profiles turned out to be an accurate
depiction of the developmental lineage and differentiation states of the
tumors. Flow cytometry has benefited from many improvements on the instrument
side, as well as the development of fluorescent antibodies and probes.
The imaging cytometers and multiplex bead assays extend even further the
power of flow cytometry by providing more information for a wider range
of samples. Armed with the right software tools to analyze the data, flow
cytometry is poised to pro vide an unprecedented close-up view of different
cell populations and their molecular components, one cell (or one bead)
at a timeref Web resources :
isolation of WBCs and platelets from whole blood : hemolysis of
RBCs
isolation of lymphocytes, monocytes and platelets from whole blood
Ficoll-Hypaque
density gradient : overlay whole blood collected with an anti-coagulant
on a Ficoll-Hypaque cushion (1:1) in a test tube. The Ficoll-Hypaque (Histopaque®)
is a polysaccharide (type II T-independent
Ags (TI-2))
with a density (1.077 g/em3) higher than the density of lymphocytes
and monocytes, but lower than the density of RBCs or granulocytes. Following
centrifugation (30' at 2,500-2,700 rpm) the pattern in the test tube will
be :
plasma layer
lymphocytes, monocytes, and platelets layer
: they can be recovered by micropipetting and then washed from Ficoll-Hypaque
with serial dilution in complete RPMI medium and adjusted to a volume of
10 × 106 cells/mL
Ficoll-Hypaque
RBCs and granulocytes will centrifuge to the
bottom of the test tube
removal of platelets by a single sucrose solution
with a density of 65%
incubation of whole blood with ADP and carbonyl
iron prior to layering on a double discontinuous gradient of
Ficoll-Hypaque selectively reduced the number of platelets and monocytes
found at the upper lymphocyte interface after centrifugation. This modification
in combination with a double discontinuous gradient results in a lymphocytes
isolation method that is simple, rapid and reproducibleref.
isolation of lymphocytes from whole blood
Polymorphprep™
Nycodenz®
OptiPrep™
Lymphoprep™(source
: Axis-Shield) is a ready made,
sterile and endotoxin tested solution for the isolation of pure lymphocyte
suspensions. The solution contains sodium diatrizoate and polysaccharide
in the following concentrations:
sodium diatrizoate 9.1% (w/v)
polysaccharide (Ficoll®) 5.7%
(w/v)
Physical-chemical characteristics:
density 1.077 ± 0.001 g/ml
osmolality 280 ± 15 mOsm
Principle of the separation technique : the
most common technique for separating leucocytes is to mix blood with a
compound which aggregates the erythrocytes, thereby increasing their sedimentation
rate. The sedimentation of leucocytes is only slightly affected and can
be collected from the upper part of the tube when the erythrocytes have
settled. Using a mixture of sodium metrizoate and polysaccharide, Bøyumref
(Bøyum, A?1968?: Separation of leucocytes from blood and bone marrow.
Scand. J. Clin Lab. Invest, 21, Suppl. 97) developed a one-step centrifugal
technique for isolation of lymphocytes. Thorsby and Bratlieref
(1970) used this technique with only slight modifications in the preparation
of pure lymphocyte suspension for cytotoxicity tests and lymphocyte cultures.
As emphasized also by other authors, Harris and Ukayioforef
(1969), Ting and Morrisref
(1971) this is a reliable, simple and quick method suitable for the preparation
of lymphocyte preparations from cadaver blood, and from anticoagulated
blood stored at room temperature for up to 6 hours.
Boyum A. Isolation of lymphocytes, granulocytes and macrophages. Scand
J Immunol 1976;(Suppl 5):9-15.
Stability and storage : Lymphoprep™ is stable
for 5 years provided the solution is kept sterile and protected from light.
Prolonged exposure to direct sunlight leads to release of iodinine from
the sodium diatrizoate molecule. This effect is negligible when working
with this solution on a
day to day basis. Lymphoprep™ should be stored at room temperature.
Separation procedure :
1. Collect blood into a tube containing anticoagulant (EDTA, heparin, ACD)
or use defibrinated blood.
2. Dilute the blood by addition of an equal volume of 0.9% NaCl.
3. Carefully layer 6 ml of the diluted blood over 3 ml Lymphoprep™ in a
12–15 mm centrifuge tube. Alternatively Lymphoprep™ can be underlayered.
Avoid mixing of blood and separation fluid. Cap the tube to prevent the
formation of aerosols.
4. Centrifuge at 800 x g for 20 minutes at room temperature (approximately
20°C) in a swing-out rotor. If the blood is stored for more than 2
hours, increase the centrifugation time to 30 minutes.
5. After centrifugation the mononuclear cells form a distinct band at the
sample/medium interface, as shown in the figure. The cells are best removed
from the interface using a Pasteur pipette without removing the upper layer.
6. Dilute the harvested fraction with 0.9% NaCl or medium to reduce the
density of the solution and pellet the cells by centrifugation for 10 minutes
at 250 x g.
Purity and viability : the described method
has been found to be rapid, simple and reliable and gives excellent results
with blood samples from most normal individuals and patients. The technique
can also be used for preparation of lymphocyte suspensions for mixed lymphocyte
culture tests. The contamination in the lymphocyte suspensions of erythrocytes
is usually between 1-5% of the total cell number. Some immature granulocytes
may follow the lymphocytes during intense immunosuppressive therapy. When
heparinized blood is used, it is essential to remove most of the platelets,
in order to avoid inhibition in the cytotoxicity test. The described washing
procedure is usually sufficient.
Biochemical
testings
fermentation tests detect fermented carbohydrate(s) and end products
of fermentation(s). As the Bacteria excrete large amounts of lactic
acid, acetic acid and formic acid, one can detect fermentation by using
a differential medium containing a pH indicator
dye.
Neisseria enzyme test
(NET) is a biochemical test that will differentiate between Neisseria
gonorrhoeae,
N.
lactamica, Neisseria
meningitidis,
and Moraxella catarrhalis.
A change of the original purple color to yellow will indicate N. meningitidis.
A blue color will indicate N. lactamica. A pink/red color resulting
from the addition of PRO reagent will indicate N. gonorrhoeae. If
there is no color change, M. catarrhalis is indicated. Optimal incubation
for this test is at a temperature of 37° C for a duration of 0.5 hours
in ambient air.
X-chromosome
inactivation pattern (XCIP) / lyonization (the random inactivation
of the X-chromosome in females that occurs early in embryogenic developmentref)
: genes located on the X-chromosome and subject to X chromosome inactivation
can be used to distinguish monoclonal from polyclonal cell populations
in
females. A prerequisite for all methods of clonality analysis based
on X-inactivation is the ability to distinguish the paternally derived
X-chromosome from the maternally derived one, based on common polymorphisms
that exist in the general population at different loci on the X-chromosomeref.
A second prerequisite is the ability to distinguish the active from the
inactive X-chromosome. Excessive skewing or clonality of hematopoiesis
is defined as :
corrected allele ratio (CR) > 3:1, which corresponds to the finding
of one allele present on the same X-inactive chromosome in > 75% of cellsref.
The allele ratio is defined as the ratio between the intensity of
PCR products from the 2 HUMARA alleles in a given sample, after digestion
with methylation-sensitive enzymes (a'/a'b', a' being the smaller allele
after digestion). The corrected ratio (CR) is defined as the allele
ratio of the precut DNA sample (a'/a'b') divided by the allele ratio of
the nonprecut sample (a/ab) specimen. This ratio compensates for potential
preferential amplification of one of the two alleles (in general, the smaller
one).
Asimakopoulos equation : percentage of clonal granulocytes (Gc)
was calculated using the following equationref,
when Rt (T-cell ratio) > Rg (granulocyte ratio); Gc: t-Rg / Rt(Rg+1) ×
100; when Rt < Rg Gc: Rg-Rt / Rg+1 × 100. This equation divides
XCIP into 3 categories:
clonal, when Gc>50% (corresponds to a Rg <0.33 in the presence of an
Rt: 1.0)
polyclonal, when Gc<50% and Rt>0.33
ambiguous, when Gc<50% and Rt<0.33
Expression of the lower allele after correcting for the degree of lyonization
in T-lymphocytes.
One potential limitation of X-inactivation assays is skewed X-inactivation
pattern (XIP), which can mimic clonal derivation of cells
costitutional excessive lyonization : females who have randomly
inactivated a preponderance of one X-chromosome (either the paternal X
or the maternal X) relative to the other, probably due to the stem cell
pool size at the time of Lyonization
acquired skewed XIP skewed XIP has been shown to increase with age,
with > 30% of the normal population having skewed XIP at 60 yearsref.
Gale et alref
found significant skewed XIP in 23% of normal females using PGK and HPRT
probes
21–31% of healthy women have a skewing pattern of X-inactivation (or lyonization)
with techniques based on DNA polymorphismref1,
ref2.
Different hypotheses have been suggested to explain the especially high
rate of excessive lyonization in hematopoietic tissue :
skewing of the X-chromosome inactivation pattern may be caused by the small
stem-cell pool size at the time of X-chromosome inactivationref.
Patterns will therefore vary from tissue to tissue. Because X-chromosome
inactivation takes place gradually over a period of time, as demonstrated
by Tan et al., this will influence the patterns obtained; e.g., inactivation
of the progenitor blood cell pool at an earlier development time would
lead to considerably more skewing in blood cells than in gut cells, which
are apparently one of the last cell populations to be inactivatedref.
Second, it is possible that somatic cell selection occurs after random
X-inactivation has been established and as a consequence of a selective
pressure on blood cellsref1,
ref2,
ref3.
Finally, stem-cell depletion or the development of clonal hematopoiesis
may also induce a nonrandom pattern of X-inactivationref.
Each of these acquired processes could result in highly skewed X-inactivation
ratios and should be associated with an increased incidence of skewing
with age. Fey et al. have reported data supporting a different incidence
in skewing between elderly and younger femalesref,
although Gale et al. were unable to demonstrate an effect of age on skewing
when the totality of published data for healthy females was considered
(n = 100)ref.
Recently, Busque et al. showed that incidence of excessive lyonization
in healthy women increased significantly with age with the HUMARA and PGK
genes. The incidence of skewing (allele ratios > 3:1) was 8.6% (14 of
162) in neonates, 16.4% (11 of 67) in women 28 to 32 years old, and 37.9%
(25 of 66) in women 60 yearsref.
A statistically significant difference in skewing rate in neonates using
DNA polymorphism (24%) compared with RNA polymorphisms (3%) has been reported.
One explanation could be that methylation was incomplete at birth on one
allele, paternal or maternal. Indeed, the opposite pattern was observed
in one elderly woman and two ET patients. In the case of the elderly woman,
the discrepancy could be due to hypermethylation of DNA on the active X-chromosome
resulting in incomplete digestion and consequently a random pattern, while
basically X-inactivation was skewed. Alternatively, the occurrence of a
mutation on one allele of the IDS gene may lead to the absence of its expression
in females with a random pattern of X-inactivation. In patients, it may
be due to hypermethylation of DNA as already described in tumorsref,
especially because of the clear clonality feature obtained by transcript
analysis in one patient (clonality in granulocytes and platelets with polyclonality
in T lymphocytes). Comparison between the DNA and RNA polymorphisms showed
some interesting results that may help clarify the interpretation of excessive
lyonization in healthy females. Both techniques showed an increase in skewing
rate between females <50 years and elderly ones. This suggests that
methylation does not increase with age, but that depletion of stem cells,
clonal selection, or the appearance of clonal hematopoiesis might occur
with age. Furthermore, using RNA polymorphisms, we confirmed the findings
of others that incidence of skewed lyonization is very low at birth. This
low incidence of excessive skewing in neonates also suggests that the estimated
number of stem cells present at the time of X-inactivation could be higher
than previously reported. These results are not in agreement with those
of other authors who found a low rate of nonrandom X-inactivation at birth
using HUMARA polymorphismref,
but the latter group of neonates was much smaller (40 vs 162)ref
technical reasons may account for some of the discrepancies in the estimation
of skewed lyonization in healthy females and possibly clonality analysis
in various malignancies. This hypothesis is supported by the low frequency
of skewed lyonization found by studying the expression of G6PD polymorphism
at the protein level compared with the high frequency found with methylation
analysis of the M27ß locusref1,
ref2,
ref3
Skewed XIP of polyclonal cells can only be distinguished from clonal cells
using
appropriate tissue controls. Because XCIP may vary from tissue to tissueref,
somatic control from embryologically related tissue (such as T lymphocytes
or buccal mucosa cells) is needed to interpret skewed patterns of X-inactivation
in BM cellsref.
For myeloid disorders in the young, T lymphocytes serve as a suitable control
XCIP, but interpretation of imbalanced XCIPs in the elderly can be difficultref.
T lymphocytes are thought to be the most suitable control for disorders
of myeloid cell origin, because they originate from the same pluripotent
stem cell, and in younger hematologically normal individuals, they have
the same XCIPref.
However, studies have also shown that there is an increased incidence of
skewing in myeloid cells of the elderlyref1,
ref2,
which may not be reflected in T-lymphocyte XCIPsref1,
ref2.
In a study, for example, an extremely skewed pattern with > 90% expression
of one allele was found in the neutrophils of 33% and T lymphocytes of
9% of hematologically normal females more than 75 years of age, but in
only 3% of younger individualsref.
In the few studies of ET patients in which T-lymphocyte XCIPs have been
compared with neutrophil patterns, the results have been conflicting. Tsukamoto
et alref
showed monoclonal XCIPs in neutrophils of all 13 patients in their study,
suggesting that ET is almost invariably a clonal disorder, although we
would only consider 6 of these patients to be assessable because of advanced
age or constitutive skewing of the T cells. In a study of 46 patients by
El-Kassar et alref,
the majority seemed to have monoclonal neutrophils (61%), but 14 patients
had polyclonal patterns in both neutrophils and T lymphocytes. Ten of these
14 patients could be studied using RNA polymorphisms. Platelet polyclonality
was confirmed in 7 patients, and monoclonal platelets were found in 3 patients,
these latter results suggesting that ET could sometimes be restricted to
the megakaryocytic line. Such lineage restriction is in marked contrast
to other clonal myeloproliferative disordersref.
DNA methylation patterns-based clonality assays (MCA) on active
and inactive X-chromosome alleles. It uses Southern
blotting
or PCR
of either ... :
restriction enzyme polymorphisms, eg :
HPRT-PCR clonality assay
phosphoglycerate kinase (PGK)-PCR
clonality assay (PPCA) : active X chromosome DNA is unmethylated at
eight CCGG sites clustered in a guanine-plus-cytosine-rich island at the
5' end of the generef
variable number tandem repeat (VNTR) sequences, eg :
M27b, a highly informative
probe detecting a polymorphic X chromosomal locus, DXS255. The polymorphism
detected at this locus is due to variable numbers of tandem repeats. The
rate of constitutional heterozygosity detected by M27b
was 88%. Normal tissue from gastrointestinal mucosa and thyroid showed
random, hence polyclonal, patterns. Nonrandom clonal X inactivation was
detected in all 22 malignant neoplasms that had been shown to be clonal
by other DNA markers, such as antigen receptor gene rearrangements or clonal
loss of heterozygosity at 17p and other loci. 16/48 normal blood leukocyte
samples (33%) showed considerably skewed X inactivation patterns. Comparison
of blood leukocytes and normal tissue indicated that in a given individual,
X inactivation patterns may be tissue specific. M27b
was used to study the clonal composition of 13 benign thyroid nodules from
12 multinodular goiters with rapid recent growth, traditionally termed
"adenomas." 9 of them were clonal, whereas 4 nodules and tissue from a
case of Graves' goiter were not, indicating that some, but not all, such
thyroid nodules may represent true clonal neoplasms. The M27b
probe permits one to study the clonal composition by the X inactivation
approach of a wide variety of solid tumors from most female patients. As
a control, normal tissue homologous to the tumor type of interest is preferable
to DNA from blood leukocytes, since the latter may show nonrandom X inactivation
patterns in a fairly high proportion of cases. M27b
may, therefore, be of limited use for the clonal analysis of neoplasms
derived from hematopoietic cellsref1,
ref2,
ref3
human androgen
receptor
assay (HUMARA)ref.
It offers several advantages over other X-inactivation-based assays because
it has a high rate of heterozygosity (>90%) as a consequence of > 20
alleles generated by the different number of trinucleotide (CAG) repeats
in the first exon at this locus in the general populationref.
The close proximity of the polymorphic trinucleotide repeat to 4 methylation
sitesref
permits a clonal analysis based on the PCRref.
One primer is labeled at the 5' end with fluorescein. The products were
analyzed and quantified with an automated DNA sequencer (ALF, Pharmacia-Biotech).
Under optimal conditions amplification efficiency of two HUMARA alleles
is near-equivalent, generating PCR products in a ratio proportional to
that of the genomic template. In contrast, reduction of template quantity,
damage of template by ultraviolet irradiation or addition of monovalent
salts (sodium chloride, sodium acetate or ammonium acetate) produces highly
variable imbalances of allelic PCR products, with a strong tendency to
preferentially amplify lower molecular weight alleles. Variability and
biasing was diminished by substitution of 7-deaza-2'-dGTP for dGTP during
amplification, an intervention which reduces stability of intramolecular
and intermolecular GC base pairingref.
This assay has been validated in the analysis of hematopoiesis and in various
hematologic disordersref1,
ref2,
ref3,
ref4
(Busque L, Maragh M, DeHart D, McGarigle C, Vose J, Armitage J, Meisinger
D, Wheeler C, Gaines L, Belanger R, Habel F, Dunbar C, Champagne M, Gross
W, Weinstein H, Antin JH, Gilliland DG: Clonality of bone marrow repopulation
after allogeneic and autologous bone marrow transplantation (ABMT). Blood
82:457a, 1993; Busque L, Gilliland DG: Clonality analysis in myelopoietic
disorders. Focus Growth Factors 5:3, 1994).
transcription-based clonality assays (TCAs) to detect mRNA polymorphismref
: only genes on the X-active chromosome are transcribedref1,
ref2,
ref3.
One advantage of this technique is the possibility to study clonality in
nonnucleated cells such as platelets and reticulocytes, which lack DNA.
Moreover, it avoids the problem of potential incomplete digestion of genomic
DNAref
glucose-6-phosphate
dehydrogenase (G6PD) : only 18% of females in the USA are informative
(heterozygous) for this chromosome with a test based on nucleotide #1311
exonic polymorphism. Detected using rtPCR-LDR
p55
/ MPP1 : a palmitoylated membrane protein with the G or T at cDNA #358
at the third exon (38% of Caucasian, African-American, and Asian
females are informative)ref.
Detected using rtPCR-LDR
iduronate-2-sulfatase
(IDS) : 41% were informativeref;
the frequency of the IDS polymorphism is 46% in Caucasian females and 39%
in African-American females; not informative in Oriental femalesref
G6PD + p55
: there is a multiple crossover between the G6PD and the p55 polymorphisms
(separated by approximately 200 kb), suggesting that these two polymorphisms
are in linkage disequilibrium; thus, approximately 50% of female
subjects are informative for clonality studies using the 2 assaysref.
Allele-specific
PCR (ASPCR)
is a simple, reproducible assay, practical for screening genomic DNA samples
for p55/G6PD genotypes, rapid clonality determination, and to determine
the linkage relationship between these closely related loci. The salient
feature of ASPCR is the performance of two PCR rounds. The first generates
template; the second, using one aliquot of first-round products in two
reaction tubes, each containing one allele-specific primer, detects each
allele. ASPCR and rtPCR-LDR produced identical p55/G6PD results in 91 normal
female genomic DNAs, and in 12 clonal hematopoietic disorder cDNAs, confirming
assay validity : 60% of females were heterozygous at one or both loci.
G6PD and p55 allelic frequencies were significantly different among African-American
men and women, but were not significantly different among Caucasian men
and women. These loci were in linkage equilibrium among African Americans,
but not among Caucasiansref.
G6PD + p55
/ MPP1 + IDS : 76%
of Caucasian females and 62% of African-American females are informative
for these assaysref
G6PD + IDS
+ p55 / MPP1 establishes the presence or absence of
clonal hematopoiesis in about 65% of female subjects
G6PD + IDS
+ p55 / MPP1 + BTK + FHL1
establishes the presence or absence of clonal hematopoiesis in about 90%
of female subjectsref
protein polymorphism :
G6PD isoenzymesref
(Fialkowref1,
ref2,
ref3)
: limited by the low frequency of heterozygotes
Clonality assays involving DNA and RNA polymorphisms are usually concordant
except in neonatesref
myeloid-specific clonality
assays :
JAK2(V617F) in polycythemia
vera (PV)
> other classic and atypical myeloproliferative disorders (MPD)ref
PCR analysis of length of CDR3 of the rearranged TCR gene, followed by
sequencing
bands other than the germ-line bands on Southern blots probed for the TcR
b gene (Vb)
anti-TCRVb monoclonal antibody staining
clonogenicity assays :
Indications :
analysis of carcinogenic transformation
effectiveness of antitumor drugs to suppress unrestricted proliferation
of cancer cells
studies of stem/progenitor cells
The conventional assays are limited to providing information about frequency
of colonies (cloning efficiency) and do not reveal the qualitative (phenotype)
attributes of individual colonies that may yield clues on mechanisms by
which cell proliferation was affected by the studied agent. Analysis
of the features of cell colonies by laser scanning cytometry (LSC)
provides a wealth of information about the progeny of individual cells.
Changes in colony size and phenotype, reflecting altered cell shape, cell
size, colony protein/DNA ratio, and expression of individual proteins,
may reveal mechanisms by which drugs suppress the proliferative capacity
of the cells. This may include inducing growth imbalance and differentiation
and modulating expression of the genes that may be associated with cell
cycle, apoptosis, or differentiation in a progeny of individual cells.
Extensions of LSC may make it applicable for automatic analysis of cloning
efficiency and multiparameter analysis of cell colonies in soft agar. Such
analyses may be useful in studies of the mechanisms and effectiveness of
antitumor drugs, in the field of carcinogenesis, and for analyzing primary
cultures and assessing tumor prognosis and drug sensitivity. The assay
can also be adapted to analysis of microbial coloniesref.
Nucleic
acid extraction
manual extraction
automated extraction : multilayer soft lithography is an alternative
to silicon-based micromachining that uses replica molding of nontraditional
elastomeric materials to fabricate stamps and microfluidic channels containing
on-off valves, switching valves, and pumps entirely out of elastomer. The
softness of these materials allows the device areas to be reduced by more
than two orders of magnitude compared with silicon-based devices. The other
advantages of soft lithography, such as rapid prototyping, ease of fabrication,
and biocompatibility, are retainedref.
from a group of cells : a series of tiny O-shaped channels, each
of which serves as a mixing vessel, is carved in a rubber postage stamp-sized
chip that drags DNA from cells. A network of inlet canals and valves feeds
each O-channel with a mixture of cells and the chemical solutions that
dilute and burst them. Tiny moving plugs in each O-channel swirl the contents
like a cement mixer, so that the cells pop and release their DNA, which
is sucked out the other side
from a single cell : for example, to identify a rare cancerous blood
cell that is about to trigger a leukaemia relapse but is swimming among
healthy neighbours. Most techniques today tend to analyse groups of cells,
which can mask a single cell's erratic behaviour. Some researchers have
started analysing single cells with conventional laboratory technique,
but shrinking the size of each chemical reaction cuts the cost of reagents
and slashes the time taken for each step
isolation of serum from the red blood cells from whole blood :
draw a volume of whole blood into a Vacutainer without using any anti-clotting
factors.
allow blood to clot at 37°C for 30 to 60 minutes.
pull the rubber stopper from the Vacutainer tubes and using a long Pasteur
pipette separate the clot from the edges of the inside of the glass.
set the tubes at 4°C overnight to allow the clot to contract.
decant the serum from the clot into a centrifuge tube (don't worry if some
red blood cells and clot material decant along with the serum).
centrifuge at 3,000 X g for 10 min to remove remaining red blood cells
and clot material.
decant the supernatant (serum) and aliquot in labeled tubes for storage.
if azide won't interfere with your applications, you should add 5-10 mM
sodium azide to this working solution tookeep bacteria from growing in
it.
there are various methods of storing prepared serum. If the serum is a
working solution one can store it at 4°C. To keep bacteria from growing
in the solution you can add sodium azide (5-10 mM final concentration)
but only if the sodium azide won't interfere with other uses of the serum.
Alternatively, serum can be stored at -80°C. Antibody will be stable
for months if the serum solution is stored at -80°C.
Imaging
following
protein dynamics in living cells
photoactivation techniques
work
by converting molecules to a fluorescent state by using a brief pulse of
high-intensity irradiation
flash photolysis of
caged compounds : caged fluorophores are fluorophores that have been
chemically modified with a caging group that quenches their fluorescenc
until a brief pulse (usually < 100 ms) of 350-nm light breaks the photolabile
bond connecting the fluorophore and the caging group
photobleaching techniques
: the photo-induced alteration of a fluorophore that extinguishes its fluorescence.
fluorescence
recovery after photobleaching (FRAP) : a region of interest (ROI) is
selectively photobleached with a high-intensity laser and the recovery
that occurs as molecules move into the bleached region is monitored over
time with low-intensity laser light. Depending on the protein studied,
fluorescence recovery can result from protein diffusion, binding/dissociation
or transport processes. Analysis of fluorescence recovery can be used to
determine the kinetic parameters of a protein, including its diffusion
constant, mobile fraction, transport rate or binding/dissociation rate
from other proteins. In experiments in which the protein of interest moves
freely, the fluorescence will recover to the initial prebleach value and
the shape of the recover can be described mathematically with a single
component recovery. Determining the effective diffusion coefficient (Deff)
and mobile fraction (Mf) of a protein from such data
is relatively straightforward, given the previous analysis of FRAP kinetics.
If the shape of the curve is complex (that is, it requires a multi-component
diffusion equation), then multiple populations of the molecule with differing
diffusion rates are present. This can occur when a molecule undergoes binding
and release from intracellular components or exists as a monomer and multimeric
forms. Alternatively, the protein might not be diffusing but might be undergoing
movement driven by molecular motors or membrane tension flow. A simple
test for determining whether a fluorescent protein moves by diffusive movement
or facilitated transport is to vary the size of the bleached area or beam
radius, w. The recovery will change with an
w2
dependece for diffusive movement only. Accurate analysis of FRAP data requires
that the bleach event is much shorter than the recovery time and preferably
as short as possible. Moreover, the recovery event must be monitored until
a recovery plateau is achieved, which is much greater than the half-time
for recovery. Images on the confocal microscope are obtained by scanning
a focused laser beam across the specimen and recording the emitted fluorescence
through a pinhole that is situated in front of the light detector. One
way to photobleach using this system is to define a region-of-interest
at the highest possible zoom, set the laser power to maximum, and set the
laser attenuation to zero. The high zoom increases the dwell time of the
laser on the bleached region per line scan (laser intensity increases proportionally
to the square of the zoom factor), which therefore greatly increases the
radiation per area. But a more advanced method is to use an acousto-optical
tunable filter (AOTF; available on more recent commercially available confocal
microscopes), which allows rapid (microsecond to millisecond) attenuation
of the laser as it scans as field. By allowing rapid switching between
the bleaching and normal beam, the AOTF allows accurate measurements of
diffusion rates in defined areas. Use of an AOTF also enables users to
photobleach virtually any pattern or shape. This allows FRAP studies to
be done on organelles of complex shapes, allowing the lateral mobility
of organelle-specific membrane and lumenal proteins to be investigated.
Selective photobleaching on a confocal microscope also provides a method
for analysing aspects of protein dynamics other than diffusion (incuding
assembly/disassembly of protein complexes in cells), the exchange of cytosolic
proteins on and off organelles, and the lifetime and fate of membrane-bound
transport intermediate. This type of analysis often requires measuring
the fluorescence signal of GFP in a specific structure or area, to compare
it with fluorescent molecules in different sites of states are known, computer
modelling can then be used to determine the parameter values (that is,
the rate constants for binding interactions and exchange times) of the
processes of interest.
Problems :
lack of recovery or partial recovery after photobleaching : possible explanations
include :
an immobile fraction of unbleached molecules in the cytoplasm that could
not diffuse into the bleached region
an immobile fraction of molecules in the bleached area that was unable
to exchange with the incoming unbleached molecules
the bleached area is not continuous with the rest of the cell (for example,
a separate membrane compartment)
reversible photobleaching of GFP : the excitationmight cause the GFP molecule
to flicker or to be sequestered in a triplet state. Both of these situations
can result in the recovery of fluorescence (in miliseconds to several seconds)
of the GFP molecule in the absence of diffusion. To control for any reversible
photobleaching of the GFP in a FRAP experiment, the FRAP conditions should
be repeated in fixed samples in which no recovery of fluorescence should
be expected. Or, alternatively, the bleach spot size could be varied and
the changes in the timescale of recovery could be confirmed
non-diffusive behaviour : measurements in FRAP studies are often complicated
by the binding and dissociation of fluorescent molecules to and from intracellular
components. This is usually reflected in the FRAP curves by the longer
recovery times, by an incomplete recovery (an immobile fraction) or by
the presence of several slopes (indicating several recovery processes over
different timescales). Kinetic modelling methods, along with computer simulations,
have been useful tools to dissect and analyse the recovery curves obtained
by FRAP. A kinetic model is characterized by biophysical parameters, such
as binding and release rate constants, diffusion constants, flow rates
and residence times. The model can be simulated on the computer for different
parameter values. Once the parameters that best fit the experimental data
have been determined, the predictions of the model can be tested experimentally
D values of the same region of interest (ROI) in the samne cell
in 2 consecutive experiments are different : a potential explanation is
damage to the photobleached area. Decreasing the bleach time, acquisition
time or the excitation beam intensity during the recovery period could
avoid damaging the cell. Using YFP rather GFP or CFP will also make it
easier to photobleach
strategies that indirectly highlight a pool of molecules by decreasing
the background fluorescence.
inverse FRAP (iFRAP) is performed
as a normal FRAP experiment with the exception that the molecules outside
a region of interest are photobleached and the loss of fluorescence from
the non-photobleached region is monitored over time. As opposed to the
rate of recovery studied using a FRAP experiment, iFRAP offers a way to
monitor the rate of movement out of a region. This method
fluorescence
localization after photobleaching (FLAP) : the same protein-of-interest
is tagged with 2 different fluorophores that co-localize when expressed
in cells. By photobleaching one of these fluorophores, a selected pool
can be highlighted and followed over time.
fluorescence
loss in photobleaching (FLIP) : a fluorescent cell is repeatedly photobleached
within a small region while the whole cell is repeatedly imaged. Any regions
of the cell that are connected to the area being bleached will gradually
lose fluorescence due to lateral movement of mobile proteins into
this area. By contrast, the fluorescence in unconnected regions will not
be affected. In addition to assessing continuity between areas of the cell,
FLIP can be used to assess whether a protein moves uniformly across a particular
cel compartment or undergoes interactions that impede its motion. Furthermore,
it can be used to reveal faint fluorescence in unconnected compartments
that normally cannot be seen against the bright fluorescence that arises
in other parts of the cell.
donor FRAP (DFRAP)
the optical voltage sensor FlaSh, made from a fusion of a GFP "reporter
domain" and a voltage-gated Shaker K+ channel "detector domain,"
has been mutagenically tuned in both the GFP reporter and channel detector
domains. The biarsenic compound FIAsH, a compound that fluoresces green
when bound to a genetically engineered protein motif (TC motif). This has
produced sensors with improved folding at 37°C, enabling use in mammalian
preparations, and yielded variants with distinct spectra, kinetics, and
voltage dependence, thus expanding the types of electrical signals that
can be detected. The optical readout of FlaSh has also been expanded from
single wavelength fluorescence intensity changes to dual wavelength measurements
based on both voltage-dependent spectral shifts and changes in FRET. Different
versions of FlaSh can now be chosen to optimize the detection of either
action potentials or synaptic potentials, to follow high versus low rates
of activity, and to best reflect electrical activity in cell types with
distinct voltages of operation
With pulses of intense laser light 10-15 s long, they can vaporize
subcellular structures inside living cells without killing them. The laser
works inside the cell without damaging the surface : the light is focused
extremely tightly, using a microscope, into a space a few hundred nm across.
atomic force
microscope (AFM) can perform keyhole surgery, adding or removing molecules
from precise locations inside a single living cell without harming it.
Using microscopic lances to remove material from fertilized eggs is now
a routine technique, but these microcapillaries are still quite clumsy
and difficult to control precisely without damaging the cell : as they
press through the cell wall, it is often deformed so badly that the cell
dies. Unlike a microcapillary, the AFM can sense the force it exerts on
the cell, making it extremely responsive. A beam of energetic ions is used
to sharpen a standard silicon AFM tip into a needle just 8 mm
long and 200 nm wide. When the researchers inserted the needle into a human
embryonic kidney (HEK) cell, the cell wall was indented by only 1 mm,
much more delicate than comparable microcapilliary procedures. The cell
membrane quickly returned to its original shape, and the needle was pushed
into the cell's nucleus. It would also be possible to monitor the chemistry
of a cell in real time by coating the needle's tip with molecules that
grab onto chemicals produced inside the cell. If the tip were designed
to fluoresce when the molecules bind, or to change its electrical charge
in a detectable way, scientists could watch the cell's response to a new
drug. The tool could be particularly useful for studying human cells in
diseases such as asthma or cystic fibrosis in which cells are often in
short supplyref
phagocytosis of molecule-loaded autologous RBCs
modified to be selectively recognized by tissue macrophages, e.g. by promoting
autologous Ig binding and C3b deposition. However, several molecules have
been shown to leak rapidly through RBC membrane due to simple diffusion.
Other molecules may be toxic to the RBC itself, thus preventing their use
as a carrier system. It is interesting to note that RBCs are "active" carrier
systems, endowed with a number of enzimatic activities that can be conveniently
explored to convert an inactive pro-drug into an active drug. This property
permits the design of a number of pro-drugs that can be synthetised with
charged chemical groups making them non-diffusible or non-toxic. Once these
chemical groups have been hydrolyzed by resident RBC enzymes, the pro-drug
is converted into an active drug that can diffuse through the RBC membrane
and thus released in circulation or at specific sites when RBC targeting
is achieved.
artificial
protocells are not truly alive, because they cannot replicate or evolve,
but they can churn out proteins for days, and could be useful for drug
production, as well as advancing the quest to build artificial life from
scratch. Vincent
Noireaux and Albert
Libchaber of the Rockefeller University in New York have managed to
package up some of the molecular machinery of a cell inside an artificial,
bacterium-sized membrane. And they can perforate the membrane with holes
that allow nutrients and energy-rich molecules to get into the cells from
the surroundings. These protocells contain all the machinery needed to
generate proteins from amino acids, so they could be used as miniature
factories, to produce proteins of industrial and medical value, routinely
produced by genetically engineered bacteria bred in fermentation vats.
But artificial cells would make much simpler protein factories, perhaps
more easily tailored to make specific products. Ready-made mixtures of
all the biomolecules that a cell needs for protein production are commercially
available, extracted from bacteria such as Escherichia coli. These
mixtures can make specific proteins, but they stop working within about
2 hours unless they are continually fed with raw materials and cleaned
of waste products. To enclose this biomolecule mixture inside membranes
like those of natural cell walls, Noireaux and Libchaber made microscopic
droplets of the cell extract suspended in oil. Soap-like molecules called
phospholipids then coated the surface of these droplets in the same way
that emulsifying agents surround the droplets in a salad dressing, stopping
them from coalescing.The researchers then coated the droplets with a second
layer of phospholipids, to form a double layer that looks just like the
membrane of a real cell. To monitor the behaviour of their cells, Noireaux
and Libchaber added DNA that encodes a fluorescent protein, so that as
the cells produce it, they start to glow. Whereas the bare cell extract
ran out of steam after 2 hours, wrapping the molecules in a membrane kept
the system 'alive' for more than twice as long. To get raw materials into
the protocells from the surroundings, the researchers added a bacterial
gene that encodes a protein called alpha-hemolysin. This protein has a
hollow, barrel-shaped structure and it inserts itself into cell membranes
to create pores. Once fitted with these molecular portals, the cells kept
churning out the fluorescent protein for days. Noireaux and Libchaber are
now working on fixing molecules to their protocell walls that will pinch
the membranes together so that the cells can divide, as bacteria do. This
could be the first step in making artificial cells that replicate. Another
big challenge is to enable the transcription and translation apparatus
to make copies of itself, so that the new cells can both function like
their 'parent'.
tissue microarrays to discover
not only the location and abundance of all human proteins, but also how
these are affected by disease states. From each cloned gene from chromosomes
X, Y, 14, and 22 (more will follow), the team produces polyhistidine-tagged
recombinant protein and uses it to generate polyclonal antibodies. In formalin-fixed
tissue samples on a slide, the proteins are denatured, unfolded. If you're
using monoclonal antibodies, you're lucky if the one antibody-binding site
is exposed. The group is producing 100–150 new recombinant proteins each
week, and five new antibodies a day. They are generating about 150 gigabytes
of data every day, and have written 60,000 lines of code in the past year
to create a database of the images and associated experimental data. Each
antibody is used to produce more than 700 pictures from different tissue
samples, amounting to 10,000 images each week. The fully automated procedure
probes tissue microarray samples from 48 normal tissues, including heart
and brain, and from the 20 cancers that occur most frequently in the Western
world, including breast, prostate, and liver.
Bibliography :
Spector, D.L., Goldman, R.D. & Leinwand, L.A. : Cells: A Laboratory
Manual, eds. Cold Spring Harbor Laboratory Press, Cold Spring Harbor,
New York, USA, 1998